QuickMix Cartridge Refill 50g Dentine:
- 1 x Refill QuickMix Cartridge
- Accessories
QuickMix Syringe Refill 10g Dentine:
- 1 x Refill QuickMix Syringe .
- Accessories
Rebilda DC – QuickMix SyringeSet :
- 1 xQuickMix Syringe10g dentine
- 1 x Futurabond DC bottle 4ml liquid
- Accessories
Direction to Use
A) Preparing the Tooth:
- Take appropriate measures to protect areas close to the pulp.
- Ensure that the tooth is properly prepared for core build-up or post luting, following standard dental procedures and techniques.
B) Bonding Material:
- Rebilda DC is used in the adhesive technique in combination with a light- or dual-curing adhesive that must be compatible with dual-curing composites.
- Consult the product’s instructions for use to determine the correct preparation (etching technique) and procedure for bonding.
C1) Using Rebilda DC in the Cartridge:
- Insert the Rebilda DC cartridge into the VOCO dispenser (type 3) or any other extruder gun with a suitable piston.
- Remove the cap, and press out the material until an equal amount has been squeezed out of both outlets.
- Insert a mixing tip type 12 and lock it with a 90degreesturn clockwise. If needed, attach an additional intraoral tip type 3 onto the mixing tip.
- For root canal application, use a mixing tip type 19 with an integrated metal cannula. You can pre-bend the cannula by hand.
- Press the lever to extrude the material and mix it automatically in the correct ratio.
C2) Using Rebilda DC in the QuickMix Syringe:
- Remove the cap and press out the material from the QuickMix syringe until an equal amount comes out of both outlets.
- Place a mixing tip type 11, ensuring the notches on the syringe and the mixing tip line up exactly. Fix the mixing tip with a 90degreesturn clockwise.
- If necessary, add an application tip type 4 on top of the mixing tip. The material is mixed automatically in the correct ratio when dispensing and can be applied directly.
D) Application of the Core Build-Up Material:
- Place a matrix around the prepared tooth to contain the material.
- Apply Rebilda DC directly from the cartridge or QuickMix syringe to the prepared tooth.
- For light-curing, use transparent matrices for optimal results. Place the light-curing device’s tip as close as possible to the surface of the core build-up and polymerize for 40 seconds using an LED/halogen lamp with a light output of at least 500 mW/cmツイ. If the distance is more than 5 mm, the curing depth may be reduced.
- For larger core build-ups, apply in layers of 2 mm and cure each layer for 40 seconds.
- After light-curing, the material can be finished immediately.
- Without light-curing, the build-up can be finished 5 minutes after its application in the mouth.